Page tree

EUDAQ integration for mulitple calorimeter by summer student Gonzalo Morras 2019

  • EUDAQ producer
  • Report


tested by Jan-Hendrik Arling and Jan Dreyling-Eschweiler, 28/02/2019

  • Lead Glass Counter number 12 from cabinet 6
  • CAEN 40 channel high voltage system, Mod. SY 127

Operation

  1. set connected channel, here for example no. 5: F0 # 4 #
  2. set current limit, here 850uA: F2 # 850 #
  3. set voltage, here  1750V: F1 # 1750 #
  4. switch on for ramping up: F10 #
  5. switch off ramping down: F11 #

IV curve for cross-check

Voltage in VCurrent in uA
500229
1000452
1500678
1750791