Page tree

Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  • Lead Glass Counter number 12 from cabinet X6
  • CAEN 40 channel high voltage system, Mod. SY 127

...

  1. set connected channel, here for example no. 5: F0 # 4 #
  2. set current limit, here 850uA: F2 # 850 #
  3. set voltage, here  1750V: F1 # 1750 #
  4. switch on for ramping up: F10 #
  5. switch off ramping down: F11 #

IV curve for cross-check

Voltage in VCurrent in uA
500229
1000452
1500678
1750791

Averaged puls at -6V trigger threshold

...