P02.1 - Public : Calculation of SDD (sample-to-detector distance)

Detector options for Rapid Access / Mail-In Users:

All values are calculated for our area detectors at a wavelength of lambda = 0.2076 Angström.

PXRD measurements - Option 1:

The area detector will be in the centered configuration and placed at a distance of 1000 mm.


Information contentQ (1/Angström)Q_min (1/Angström)2Theta at 0.2076 Angströmcomparable 2Theta in CuKalpha1 radiation
Full Debye-Scherrer ring~ 6.1~ 0.1~ 11.59°~ 97°
Corner position on detector~ 8.5~ 0.1~ 16.1°-


Example file for LaB6 standard measured on VAREX detector: LaB6_detpos1_SDD1000center_60sec-00050_00000.xy

PXRD measurements - Option 2:

The area detector will be off-centered to the corner position of the detector and placed at a distance of 2300 mm.

Please be aware that in this configuration due to the large distance, air scattering will become more severe.

Information contentQ (1/Angström)Q_min (1/Angström)2Theta at 0.2076 Angströmcomparable 2Theta in CuKalpha1 radiation
Quarter Debye-Scherrer ring~ 5.3~ 0.1~ 10.1°~ 81.5°
Corner position on detector~ 7.6~ 0.1~ 14.1°~ 132°


Example file for LaB6 standard measured on VAREX detector: LaB6_SDD2300_10s-00023_00000.xy

Total Scattering/PDF measurements - Option 3:

The area detector will be in the centered configuration and placed at a distance of 280 mm.

Information contentQ (1/Angström)Q_min (1/Angström)2Theta at 0.2076 Angströmcomparable 2Theta in CuKalpha1 radiation
Quarter Debye-Scherrer ring~ 18.8~ 0.25~ 36.2°-
Corner position on detector~ 23.6~ 0.25~ 45.9°-


Example file for LaB6 standard measured on VAREX detector: LaB6_detpos3_SDD280center_300sec_sum_4921_4925_00000.xy

Total Scattering/PDF measurements - Option 4:

The area detector will be off-centered to the corner position of the detector and placed at a distance of 280 mm.

Information contentQ (1/Angström)Q_min (1/Angström)2Theta at 0.2076 Angströmcomparable 2Theta in CuKalpha1 radiation
Quarter Debye-Scherrer ring~ 28.25~ 0.25~ 55.6-
Corner position on detector~ 32.15~ 0.25~ 64.2°-


Example file for LaB6 standard measured on VAREX detector: LaB6_detpos4_SDD280corner_300sec_sum_5571_5575_00000.xy

Attachments:

Pos1.jpg (image/jpeg)
Pos3.jpg (image/jpeg)
LaB6_SDD2300_10s-00023_00000.xy (application/octet-stream)
LaB6_detpos4_SDD280corner_300sec_sum_5571_5575_00000.xy (application/octet-stream)
LaB6_detpos1_SDD1000center_60sec-00050_00000.xy (application/octet-stream)
LaB6_detpos3_SDD280center_300sec_sum_4921_4925_00000.xy (application/octet-stream)